FTIR QUANTITATIVE ANALYSIS OF LOW CONCENTRATION FURALDEHYDE


Zheng Zegen,Zhang Yuqi Analytical and Testing Center

DOI:10.11835/j.issn.1674-4764.1991.02.004

Received ,Revised , Accepted , Available online July 01, 2015

Volume ,1991,Pages -

  • Abstract
This paper studies the determination methods offuraldehyde and compares them with each other in a furaldehyde-benzenesample system that contains less than 10% of furaldehyde.The recentChinese national standard method is that of hydroxylamine(GB 1926-80).This method has an error approximately greater than 5%,and it has moresteps and is time-consuming.The Nicolet DX quantitative analysis packagecauses even greater error,while FTIR method has errors less than 5%by selecting proper baselines and drawing a work curve.FTIR methodis simple,quick and nondestructive,and needs only few amounts ofsamples.It's suitable to the samples of the system of the experiment.